Field/Laboratory/Industrial
Instruments, Controllers & Sensors
Multi-parameter Bench Meters
Please click above to view this range
pH, Reference, Conductivity, Redox,
Dissolved Oxygen, Ion Selective Electrodes, Calibration
Solutions,
Hollow Cathode
Lamps
and
Deuterium
Lamps
|
Process
Control Instruments : Resistivity
: Industrial Online Resistivity
Monitor /
Controller for
pure, ultrapure, reagent grade water and batches of incoming process
water.
Our Alpha
RES1000 Resistivity controller is ideal for pure, ultrapure, and
reagent grade water and batches of incoming process water. The meter
can be customised by the user
through it's advanced setup menu offering flexibility in matching
the controller's functions to suit individual's specific requirements.
The user has the ability to measure two
Selectable Resistivity Measurement Ranges in one controller via its
IP54 front panel. High-level accuracy of ±1% of full scale can be
obtained with appropriate cells and correct temperature coefficient
some of the features offered from the Alpha RES1000 include : Choice
of Cell Constant (0.01, 0.1) for accurate control in any solution,
Hold Relay for use with float switches/flow switches and other
controllers as a failsafe function, Two Level Password Protection
prevents un authorized tampering with settings, 0 to 2000 second
time delay adjustment on control and alarm delays and the choice of
two Galvanically Isolated Scaleable 0-20/4-20 mA Outputs for use
with your SCADA or other control devices.
Applications for the Alpha RES1000 include :
General :
Use for virtually any batch or online applications
where rapid, accurate control is required. Great for OEM/system integrator.
Water Purification/Treatment :
Suitable for pure, ultrapure, and reagent grade
water batches of incoming process water.
Regulatory :
Hook to recorder to document data for regulatory
compliance.
Main Features :
- Two Selectable Resistivity Measurement Ranges in one
controller via its IP54 front panel.
- High-level accuracy of ±1% of full scale can be obtained with
appropriate cells and correct temperature coefficient
- User-Customization through Advanced Setup Menu offers
flexibility in matching the controller's functions to suit
individual's specific requirement
- Choice of Cell Constant (0.01, 0.1) for accurate control in
any solution
- Hold Relay for use with float switches/flow switches and other
controllers as a failsafe function
- Two Level Password Protection prevents un authorized tampering
with settings0 to 2000 Second Time Delay Adjustment on control
and alarm delays
- Two Galvanically Isolated Scalebale 0-20/4-20 mA Outputs
- Wash Contact Relay controls electrodes cleaning systems at
desired duration and frequency
- Adjustable Hysteresis (Dead Band) prevents rapid contact
switching near set point
- Non-Volatile Memory retains all stored parameters and
calibration data even if power fails
- Line Resistance Compensation against intrinsic cable
resistance for longer cable connection
- Large Dual Display shows measurement with temperature
simultaneously - features clear multiple icons, set points, and
status messages
- Choice of Temperature Sensor Pt100/Pt1000 with 2-wire or
3-wire temperature input selection
- Easy Installation and Wiring with detachable plug-in
connectors
SPECIFICATIONS :
| Product
Code |
EC-RESCTP1001/2 |
| Resistivity Range |
0.000-1.999;
0.00-19.99 Mohm |
| Resolution &
Accuracy |
0.001;
0.10 Mohm & 1% Full Scale |
| Temperature Range |
-9.9
to 125.0 °C |
| Resolution &
Accuracy |
0.1 °C
& ± 0.5 °C |
| Temperature
Compensation |
Ultra-pure
water or linear (0.00 to 10.00%) |
| Type of Controller
Setting |
Limit |
| Pickup/Dropout Delay |
0 to
2000 seconds |
No. of Relays
Maximum Voltage
Maximum Current
High/Low Hysteresis Band |
2
SPDT; 1 Alarm
250 VAC
3A;
0 to 10% Full Scale |
| Contact Set-points |
Lo/Lo;
Lo/Hi; Hi/Lo; Hi/Hi (selectable) |
| Alarm Function |
Latching
/ Pulse (software) |
| Power |
120 or
240 VAC, 50/60 Hz |
| Dimensions |
1/4
DIN (96 x 96 mm) |
Click