Specification
|
Linearisation: |
Included in the software, and available on any of the 5 elements. Achievable on both single and Multi Point calibrations. |
| Aspiration Rate: |
3-5.5ml/min |
| Optimal Range: |
| Single Point Calibration: |
|
Multi Point Calibration: |
| Na |
0.05
|
- 60ppm |
|
Na |
0.05
|
- 1000ppm |
| K |
0.05
|
- 100ppm |
|
K |
0.05
|
- 1000ppm |
| Li |
0.1
|
- 50ppm |
|
Li |
0.1
|
- 1000ppm |
| Ca |
2.5
|
- 100ppm |
|
Ca |
2.5
|
- 1000ppm |
| |
|
Ba |
30
|
- 3000ppm |
|
| Reproducibility: |
<1% coefficient of variability for 20 consecutive samples over 10 minutes (after instrument stabilisation) at concentrations of 100ppm or less. |
| Limits of Detection: |
| Na |
|
0.02ppm. |
| K |
|
0.02ppm. |
| Li |
|
0.05ppm. |
| Ca |
|
1.0ppm. |
| Ba |
|
10ppm. |
|
| Time to Stability: |
Less than 15 seconds after sample is introduced into the flame. |
| Drift: |
Less than 1% per 30 minutes after instrument stabilisation. |
| Specifity: |
Na/K/Li = <0.5% to each other when equal in concentration at <100ppm. |
| Number of Parameters Measured: |
Simultaneous display of Na, K, Ca, Li, Ba. |
| Size: |
51 cm high x 38 cm wide x 41 cm deep. |
| Weight: |
Instrument = 14 kg
Shipping = 21 kg |
|
|